ty -jour a2 -latif,faisal au -krishnan,prakash au -tarricone,arthur au -gee -gee -gee,allen au -farhan,serdar au -kamran -kamran,haroon au -kini -kini -kini,annapoorna au -anapoorna au -sharma -sharma,sharma,samin py -20221 da -20221 da -20221/06/02 TI-在表面股动脉病变中NAV6栓塞滤波器和蜘蛛晶体EPD滤波器之间的比较和分析SP -9047596 VL -2021 AB-
客观的。为了比较在PAD患者的下肢动脉切除术中产生的栓塞碎片中,SpiderFX EPD和Emboshield NAV6滤波器之间的安全性和功效。
材料和方法。在2014年1月至2015年10月之间,通过定向动脉切除术(Silverhawk),旋转动脉切除术(JetStream)或激光动脉粥样硬化切除术(Turbo Elite)治疗507例有症状性周围动脉疾病的患者。Emboshield Nav6(
n = 161) and SpiderFX (
n = 346) embolic protection devices were used with each of the 3 atherectomy devices. The primary study endpoint was 30-day freedom from major adverse events (MAEs). An MAE was defined as death, MI, TVR, thrombosis, dissection, distal embolization, perforation at the level of the filter, and unplanned amputation. A descriptive comparison of the MAE rates between Emboshield NAV6 and SpiderFX embolic protection devices was conducted.
结果。接受Emboshield NAV6滤波器的患者中,从重大不良事件(MAE)率的自由度为92.0%(CI:86.7%,95.7%),而接受SpiderFX过滤器的患者中的自由度为91.6%(CI:88.2%,94.3%)。((
p
=
0.434
)。在Emboshield NAV6组中,从重大不良事件率的自由度为86.7%的下限高于83%的性能目标(
p
<
0.0008
)。
结论。在下肢治疗中,Emboshield NAV6和SpiderFX EPD过滤器之间没有显着的临床结果差异。该评估表明使用任何一种过滤器设备来治疗患有下肢病变的PAD患者的安全性和功效。SN -0896-4327 UR -https://doi.org/10.1155/2021/9047596 do -10.1155/2021/9047596 JF-介入性心脏病学杂志 - Hindawi kw -er -er- er- er- er- er- er- er- er--